28.02.2013 Views

The Principles of Clinical Cytogenetics - Extra Materials - Springer

The Principles of Clinical Cytogenetics - Extra Materials - Springer

The Principles of Clinical Cytogenetics - Extra Materials - Springer

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Fundamentals <strong>of</strong> Microscopy 85<br />

Immersion Oil<br />

As light rays enter and exit a lens, they can reflect (refract) <strong>of</strong>f <strong>of</strong> its outer and inner surfaces. This<br />

occurs because the lens and air each have a different refractive property (refractive index).<br />

Most microscopes are engineered to reduce the incidence refraction to that occurring between the<br />

condenser and microscope slide and between the specimen and objective lens. Immersion oil can be<br />

used to greatly reduce the remaining refraction by removing the air from these spaces, filling them<br />

with a substance that has a nearly identical refractive index. As a result, more light rays will exit one<br />

surface and pass into the next than would occur if air occupied these spaces, thus increasing the<br />

resolving capacity <strong>of</strong> the microscope.<br />

A question that is <strong>of</strong>ten raised is whether oiling a condenser will actually improve the image<br />

quality to the extent that it is worth having to clean an oiled condenser. <strong>The</strong> highest operating NA <strong>of</strong><br />

an unoiled condenser and objective lens is slightly less than 1.0, as the refractive index <strong>of</strong> the air (1.0)<br />

prevents a higher operating NA (3). If the NA <strong>of</strong> a condenser or objective lens is less than 1.0, no<br />

benefit results from applying immersion oil.<br />

Even with higher NAs, the loss might not be <strong>of</strong> practical significance. Remembering that resolving<br />

capacity must be balanced with image contrast, the act <strong>of</strong> applying immersion oil increases the<br />

operating numerical aperture, but also lowers image contrast. In turn, closing the aperture diaphragm<br />

improves image contrast but also lowers the resolving capacity <strong>of</strong> the microscope. Thus, the application<br />

<strong>of</strong> immersion oil to a condenser might only be beneficial when a specimen can be viewed at a<br />

reasonable contrast while the aperture diaphragm <strong>of</strong> the condenser is nearly wide open.<br />

Immersion oil comes in several types that vary in viscosity and fluorescent properties. Each is<br />

formulated to have a refractive index <strong>of</strong> 1.5150 ± 0.0002 at 23°C (essentially the refractive index <strong>of</strong><br />

glass).<br />

Non-cover-slipped microscope slides should be cleaned <strong>of</strong> immersion oil at the completion <strong>of</strong><br />

microscopic analysis. Immersion oil can cause fading <strong>of</strong> unprotected G-banded chromosomes if left<br />

on the slides for prolonged periods <strong>of</strong> time.<br />

<strong>The</strong> Microscope Stage and Coordinate Location<br />

<strong>The</strong> microscope stage provides a flat, level surface for the microscope slide and a means <strong>of</strong> affixing<br />

the slide to the stage. Controls on a mechanical stage allow the microscope slide to be moved<br />

along x- and y-axes. Mechanical stages usually have a coordinate grid on each axis to precisely identify<br />

the location <strong>of</strong> an object on the slide. <strong>The</strong> microscope stage can also be moved in an up-and-down<br />

manner (z-axis) by using the coarse and fine focus controls.<br />

Coordinate Location<br />

Recording accurate coordinates is essential for documentation <strong>of</strong> cytogenetic findings. In most<br />

instances, notation <strong>of</strong> the x and y coordinates are used for this purpose.<br />

VERNIER GRIDS AND ENGLAND FINDERS ®<br />

When a metaphase is to be relocated at a microscope other than that used for the original analysis,<br />

a system <strong>of</strong> coordinate conversion between the two microscopes needs to be employed.<br />

Microscopes <strong>of</strong> the same manufacturer and model can <strong>of</strong>ten have their stages aligned so that the<br />

coordinates <strong>of</strong> one scope can be used at another. Vernier grids or England Finders ® allow for easy<br />

conversion <strong>of</strong> coordinates between similar microscopes whose stages cannot be aligned or when the<br />

microscopes are made by different manufacturers. This technique provides a printed grid whose value<br />

is read at one microscope and then simply relocated at the second.<br />

Microscope Slides, Cover Slips, and Mounting Media<br />

<strong>The</strong> microscope slides, cover slips, and mounting media play a significant role in the contrast and<br />

resolution <strong>of</strong> an image. Microscope slides and cover slips should be made from high-quality glass to

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!