<strong>Real</strong> <strong>Ear</strong> Measurements 28<br />

Figure 33 – Illustrating <strong>the</strong> importance of <strong>the</strong> correct placement of <strong>the</strong> probe tube (Dirks<br />

<strong>and</strong> Kincaid, 1987). Think of <strong>the</strong> abscissa as <strong>the</strong> length of <strong>the</strong> ear canal (3.0 cm = 30<br />

mm) <strong>and</strong> what happens <strong>to</strong> measured SPL near <strong>the</strong> TM as <strong>the</strong> probe tube is placed<br />

along this line. For<br />

example, at 2000 Hz<br />

<strong>the</strong> measured SPL<br />

@ 30 mm from <strong>the</strong><br />

TM is ~ 2 dB less<br />

than measured at<br />

<strong>the</strong> TM! Thus, for<br />

2000 Hz <strong>and</strong> below<br />

<strong>the</strong> meaured SPL<br />

would be essentially<br />

<strong>the</strong> same regardless<br />

of where <strong>the</strong> probe<br />

tube is placed along<br />

<strong>the</strong> ear canal. On<br />

<strong>the</strong> o<strong>the</strong>r h<strong>and</strong>, @ 8000, if <strong>the</strong> clinician placed <strong>the</strong> end of <strong>the</strong> probe tube ~12 mm from<br />

<strong>the</strong> TM, <strong>the</strong> measured SPL would be ~ 12 dB less than what is present at <strong>the</strong> TM. The<br />

o<strong>the</strong>r curves @ 3000-<strong>7000</strong> Hz reveal similar findings.<br />

Figure 34 – Illustrating <strong>the</strong> proper placement of <strong>the</strong> probe tube in <strong>the</strong> ear canal where<br />

<strong>the</strong> measured REUG @ 6000 Hz is @ 0 dB gain. This is ano<strong>the</strong>r method <strong>to</strong> verify that<br />

<strong>the</strong> probe tube is in its proper place along <strong>the</strong> ear canal.