<strong>Real</strong> <strong>Ear</strong> Measurements 51<br />

Measuring if <strong>the</strong> REIG of a BICROS fitting matches a valid prescriptive<br />

target<br />

Figure 65 – Measuring overall REIG<br />

<strong>to</strong> target for a BICROS aid.<br />

2.11 <strong>Using</strong> <strong>the</strong> Frye <strong>7000</strong> as a spectrum analyzer <strong>to</strong> measure feedback.<br />

STEP-BY-STEP<br />

1. From <strong>the</strong> Opening screen, press [F2] <strong>to</strong> enter <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Navigation screen.<br />

2. Press [F4] <strong>to</strong> enter <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> SPL<br />

screen.<br />

3. Press [MENU] <strong>to</strong> enter <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> SPL<br />

Menu.<br />

NOTE: Use <strong>the</strong> UP/DOWN arrows <strong>to</strong> select a<br />

menu option <strong>and</strong> press <strong>the</strong> RIGHT arrow <strong>to</strong><br />

change a selection.<br />

a. Select Graph as <strong>the</strong> Data/Graph<br />

option under <strong>the</strong> Display<br />

heading.<br />

b. Turn <strong>the</strong> Reference Mic <strong>to</strong> Off<br />

under <strong>the</strong> Measurements<br />

heading.<br />

c. Press [EXIT] <strong>to</strong> return <strong>to</strong> <strong>the</strong> test<br />

screen.<br />

First Measure: REIG for better ear<br />

a. Reference microphone enabled <strong>and</strong> placed over <strong>the</strong><br />

poor ear<br />

b. Probe microphone in <strong>the</strong> good ear<br />

c. Level <strong>and</strong> s<strong>to</strong>re as UNAIDED<br />

Second Measure: REAG for better ear<br />

a. Aid in good ear <strong>and</strong> both microphones are ON<br />

b. S<strong>to</strong>re as AIDED<br />

Difference REIG curve shows benefit from BICROS aid <strong>to</strong><br />

target<br />

Figure 66 – <strong>Real</strong> <strong>Ear</strong> SPL Menu