Using the FONIX 7000 to Verify Coupler and Real-Ear Performance ...
Using the FONIX 7000 to Verify Coupler and Real-Ear Performance ...
Using the FONIX 7000 to Verify Coupler and Real-Ear Performance ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Real</strong> <strong>Ear</strong> Measurements 39<br />
Figure 50 – <strong>Real</strong> ear navigation screen function keys<br />
3. Press [MENU] <strong>to</strong> open <strong>the</strong> real ear audiogram menu <strong>and</strong> select <strong>the</strong> appropriate<br />
settings based on <strong>the</strong> patient’s audiogram <strong>and</strong> <strong>the</strong> hearing aid parameters.<br />
These settings are important for <strong>the</strong> generation of real ear targets.<br />
4. The clinician may now enter <strong>the</strong> patient’s audiogram. Use [F1] <strong>to</strong> select an ear<br />
<strong>and</strong> press [F2] <strong>to</strong> select HTL <strong>to</strong> enter <strong>the</strong> patient’s hearing threshold levels.<br />
Use <strong>the</strong> UP/DOWN arrow <strong>to</strong> select a frequency <strong>and</strong> use <strong>the</strong> RIGHT/LEFT arrows <strong>to</strong> select <strong>the</strong><br />
threshold level in 5 dB steps.<br />
Figure 51 – Audiogram entry screen<br />
5. If a patient’s hearing loss is conductive or mixed, it is important <strong>to</strong> enter bone<br />
conduction thresholds when using <strong>the</strong> NAL-NL1 target. Press [F2] <strong>to</strong> select Bone<br />
<strong>and</strong> use <strong>the</strong> procedure listed in Step 4 above <strong>to</strong> enter <strong>the</strong> patient’s bone<br />
conduction thresholds.<br />
6. Press [F2] <strong>to</strong> select LDL <strong>to</strong> enter a patient’s measured LDLs using <strong>the</strong> same<br />
procedure listed in Step 4 above or press [F3] <strong>to</strong> use predicted LDLs.<br />
7. If performing a binaural fitting <strong>and</strong> a patient has a similar hearing loss bilaterally,<br />
you may press [F4] <strong>and</strong> select COPY EAR <strong>to</strong> copy data <strong>to</strong> <strong>the</strong> opposite ear. To<br />
change individual threshold values as necessary, press [F1] <strong>to</strong> select <strong>the</strong><br />
appropriate ear <strong>and</strong> repeat Step 4 in order <strong>to</strong> make changes <strong>to</strong> <strong>the</strong> copied data.<br />
8. Press [EXIT] <strong>to</strong> return <strong>to</strong> <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Navigation screen.