<strong>Real</strong> <strong>Ear</strong> Measurements 39<br />

Figure 50 – <strong>Real</strong> ear navigation screen function keys<br />

3. Press [MENU] <strong>to</strong> open <strong>the</strong> real ear audiogram menu <strong>and</strong> select <strong>the</strong> appropriate<br />

settings based on <strong>the</strong> patient’s audiogram <strong>and</strong> <strong>the</strong> hearing aid parameters.<br />

These settings are important for <strong>the</strong> generation of real ear targets.<br />

4. The clinician may now enter <strong>the</strong> patient’s audiogram. Use [F1] <strong>to</strong> select an ear<br />

<strong>and</strong> press [F2] <strong>to</strong> select HTL <strong>to</strong> enter <strong>the</strong> patient’s hearing threshold levels.<br />

Use <strong>the</strong> UP/DOWN arrow <strong>to</strong> select a frequency <strong>and</strong> use <strong>the</strong> RIGHT/LEFT arrows <strong>to</strong> select <strong>the</strong><br />

threshold level in 5 dB steps.<br />

Figure 51 – Audiogram entry screen<br />

5. If a patient’s hearing loss is conductive or mixed, it is important <strong>to</strong> enter bone<br />

conduction thresholds when using <strong>the</strong> NAL-NL1 target. Press [F2] <strong>to</strong> select Bone<br />

<strong>and</strong> use <strong>the</strong> procedure listed in Step 4 above <strong>to</strong> enter <strong>the</strong> patient’s bone<br />

conduction thresholds.<br />

6. Press [F2] <strong>to</strong> select LDL <strong>to</strong> enter a patient’s measured LDLs using <strong>the</strong> same<br />

procedure listed in Step 4 above or press [F3] <strong>to</strong> use predicted LDLs.<br />

7. If performing a binaural fitting <strong>and</strong> a patient has a similar hearing loss bilaterally,<br />

you may press [F4] <strong>and</strong> select COPY EAR <strong>to</strong> copy data <strong>to</strong> <strong>the</strong> opposite ear. To<br />

change individual threshold values as necessary, press [F1] <strong>to</strong> select <strong>the</strong><br />

appropriate ear <strong>and</strong> repeat Step 4 in order <strong>to</strong> make changes <strong>to</strong> <strong>the</strong> copied data.<br />

8. Press [EXIT] <strong>to</strong> return <strong>to</strong> <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Navigation screen.