<strong>Coupler</strong> Measurements 5<br />
<strong>the</strong> CIC coupler <strong>and</strong> <strong>to</strong> activate <strong>the</strong> CIC sub-menu (Press [F1] “<strong>Coupler</strong>”, <strong>the</strong>n [Menu],<br />
scroll down using <strong>the</strong> down arrow key <strong>to</strong> “<strong>Coupler</strong> type” <strong>and</strong> <strong>the</strong>n <strong>the</strong> right arrow key <strong>to</strong><br />
“CIC”. Then press [Exit]). Finally, Figure 7 <strong>and</strong> Figure 8 illustrate <strong>the</strong> new nonst<strong>and</strong>ardized<br />
open coupler <strong>and</strong> open BTE coupled <strong>to</strong> <strong>the</strong> open coupler. Figure 9<br />
reports <strong>the</strong> frequency response of this hearing aid measured in <strong>the</strong> open coupler (Curve<br />
1), HA-1 coupler (Curve 2), HA-2 coupler (Curve 3) <strong>and</strong> unaided KEMAR (Curve 4).<br />
Although <strong>the</strong> frequency response of <strong>the</strong> open coupler does not superimpose upon <strong>the</strong><br />
unaided KEMAR response, <strong>the</strong> curve measured using <strong>the</strong> open coupler is closer <strong>to</strong> <strong>the</strong><br />
unaided KEMAR response than when measured in ei<strong>the</strong>r <strong>the</strong> HA-1 or HA-2 coupler.<br />
Battery type<br />
Selection<br />
But<strong>to</strong>ns<br />
Battery pill<br />
HA-2 <strong>Coupler</strong><br />
Microphone must be<br />
in <strong>the</strong> center <strong>and</strong><br />
facing <strong>to</strong> <strong>the</strong> right<br />
Figure 1 – Configuration of <strong>the</strong> hearing aid test chamber for performing <strong>the</strong><br />
electroacoustic analysis of a BTE hearing aid.