<strong>Real</strong> <strong>Ear</strong> Measurements 38<br />

L, M. H: <strong>the</strong> real<br />

ear targets<br />

15. Use [F2] <strong>to</strong> select REAR 3, which uses a 65 dB SPL Speech signal.<br />

16. Repeat Steps 13 <strong>and</strong> 14 above <strong>to</strong> measure <strong>the</strong> output of <strong>the</strong> hearing aid <strong>to</strong> an<br />

average input.<br />

17. Use [F2] <strong>to</strong> select REAR 4, which uses an 80 dB SPL DigSpeech signal.<br />

18. Repeat Steps 13 <strong>and</strong> 14 above <strong>to</strong> measure <strong>the</strong> output of <strong>the</strong> hearing aid <strong>to</strong> a loud<br />

input.<br />

19. Use [F2] <strong>to</strong> select REAR 5, which uses a 90 dB SPL pure-<strong>to</strong>ne sweep.<br />

20. Press [START] <strong>to</strong> begin <strong>the</strong> measurement. The test will s<strong>to</strong>p au<strong>to</strong>matically.<br />

Compare this curve, which is <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Saturation Response (RESR90), <strong>to</strong><br />

<strong>the</strong> patient’s LDLs (labeled “U”) <strong>and</strong> ensure that <strong>the</strong> curve never exceeds <strong>the</strong><br />

patient’s LDLs.<br />

2.5 Performing live speech mapping <strong>to</strong> verify <strong>the</strong> output, in real-time, of a hearing aid<br />

in response <strong>to</strong> an actual speech signal. This <strong>to</strong>ol is best used as a counseling<br />

<strong>to</strong>ol for patients <strong>and</strong> <strong>the</strong>ir family members.<br />

l, m, h: unamplfied<br />

LTASS<br />

STEP-BY-STEP<br />

LDL<br />

Threshold<br />

Figure 49 – Visible Speech unaided response<br />

1. From <strong>the</strong> Opening screen, press [F2] <strong>to</strong> enter <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Navigation screen.<br />

2. Press [F1] in <strong>the</strong> <strong>Real</strong> <strong>Ear</strong> Navigation screen <strong>to</strong> enter <strong>the</strong> audiogram entry<br />

screen.