On an iH600 or iH1000 system, the following alarms will cause only the HCMB pump to shut<br />

down, whether 'run til crash' is selected or not:<br />

• HCMB CURRENT HIGH/LOW * (error numbers 710 and 712).<br />

• HCMB THERMISTOR HIGH (error number 912).<br />

• - (error number 701) ✝ .<br />

5.12Emergency stop<br />

Notes:<br />

The emergency stop switch is not an electrical isolator, unless the disconnect box accessory is<br />

fitted (See Section 5.17).<br />

The use of emergency stop will not automatically shut down the load lock pump (if fitted); you<br />

must shut down the load lock pump separately.<br />

To shut down the iH system in an emergency, press the emergency stop switch (Figure 1-7,<br />

item 1) on the dashboard. Alternatively, you can operate the emergency stop controls in your<br />

own control system if you have connected your emergency stop circuit to the iH system as<br />

described in Section 3.15.<br />

When emergency stop is selected:<br />

• The HCDP pump (and the HCMB pump, on iH600 and iH1000 systems) is switched off.<br />

• The solenoid-valve(s) in the Gas Module close, to switch off the supply of nitrogen to the<br />

HCDP pump.<br />

• The Pump Display Terminal will display 'STOP ACTIVATED'.<br />

• The LED on the On button on the Pump Display Terminal (Figure 1-8, item 1) will go off.<br />

5.13 Restart the HCMB pump after automatic pump shut-down<br />

(iH600/iH1000 only)<br />

WARNING<br />

If a pump is automatically shut down, ensure that it is safe to restart before<br />

you restart the pump<br />

If the HCMB pump has been automatically shut down due to an alarm (see Section 5.11), use<br />

the Switch On/Booster Pump menu to restart the HCMB pump after the alarm condition has<br />

cleared.<br />