Use the following procedure to connect the iH system to your exhaust-extraction system:<br />

refer to Figure 1-1.<br />

1. If the check valve (18) is not required on your application:<br />

• remove the exhaust enclosure (26) by removingthe screws connectingto the rear<br />

bulkhead cover and exhaust enclosure top plate.<br />

• undo and remove the clamp and ‘O’ ring(22 and 21) which secures the check valve to the<br />

elbow (20).<br />

• retain the check valve for possible future use.<br />

2. If neither the check valve (18) nor elbow (20) is required on your application, or if the elbow is<br />

required in a different orientation:<br />

• remove the exhaust enclosure (26) by removingthe screws connectingto the rear<br />

bulkhead cover and exhaust enclosure top plate.<br />

• remove the clamps insulation jacket (24) and retain.<br />

• undo, or remove, the clamped ‘O’ ring that secures the elbow to the exhaust pipe (19).<br />

• reposition, or remove, the elbow (retain both elbow and check valve for possible future<br />

use if removed).<br />

3. Remove the temporary cover or blankingplate from the exhaust outlet. Retain blanking<br />

plate for future use. Retain temporary cover for future use on non-contaminated pumps<br />

only. Use the NW40 trapped 'O' ringand clamp supplied to connect the exhaust outlet on the<br />

check valve (18), or elbow (20), or exhaust pipe (19), to your exhaust extraction system.<br />

4. If the check valve was removed or elbow was removed or reorientated:<br />

• refit the clamps insulation jacket (if removed) on the NW40 joint to your exhaust<br />

extraction system. Use Figure 6-3 as a guide.<br />

• refit the exhaust enclosure (if possible) usingthe screws connectingto the rear bulkhead<br />

cover and exhaust enclosure top plate.<br />

3.11 Connect to your factory extraction system (optional)<br />

As supplied, the temperature control system fitted is sufficient to maintain the iH system at<br />

the correct operating temperature; you do not need to connect an air-extraction system to the<br />

iH system. However, if required you can connect your factory extraction system to the<br />

air-extraction port on the top of the main enclosure (see Figure 2-1).<br />

Note:<br />

The iH system is supplied with a finger guard punched into the top enclosure.<br />

iH Dry Pumping Systems 3-15