Warning message Advisory message Meaning/recommended action<br />
WARNING 101 (a) there is a fault in the Control SMA, or<br />
(b) the emergency stop button has been<br />
pressed to shut down the iH system, or<br />
(c) there has been a transient electrical supply<br />
failure (longer than the hold-up time of the iH<br />
power supply).<br />
WARNING 113 There is a fault in the ADC (analogue-to-digital<br />
convertor) in the Control SMA; the Control<br />
SMA must be replaced.<br />
WARNING 209 SEE MANUAL The electrical supply voltage is too low. Check<br />
the electrical supply and the connections to the<br />
iH system and rectify as necessary.<br />
WARNING 211 SEE MANUAL The electrical supply voltage is too high.<br />
Check the electrical supply and the<br />
connections to the iH system and rectify as<br />
necessary.<br />
WARNING 309<br />
DP CURRENT LOW<br />
WARNING 311<br />
DP CURRENT HIGH<br />
WARNING 409<br />
DP POWER LOW<br />
WARNING 411<br />
DP POWER HIGH<br />
WARNING 412<br />
DP POWER HIGH<br />
OPEN CIRCUIT<br />
CHECK CONNECTION<br />
LEAK/DEBRIS<br />
SERVICE PUMP<br />
OPEN CIRCUIT<br />
CHECK CONNECTION<br />
LEAK/DEBRIS<br />
SERVICE PUMP<br />
LEAK/DEBRIS<br />
SERVICE PUMP<br />
The electrical current drawn by the HCDP<br />
pump is less than normal. Check that the<br />
electrical connections are correct and rectify as<br />
necessary. If the electrical connections are<br />
correct, a fuse in the Electrics Box may have<br />
failed.<br />
The electrical current drawn by the HCDP<br />
pump is higher than normal. There may be a<br />
leak in the system or process debris in the<br />
pump, or the pump motor may have failed.<br />
Inspect the pipelines and service the pump.<br />
The electrical power consumption of the<br />
HCDP pump is less than normal. Check that<br />
the electrical connections are correct and<br />
rectify as necessary. If the electrical<br />
connections are correct, a fuse in the Electrics<br />
Box may have failed.<br />
The electrical power consumption of the<br />
HCDP pump is higher than normal. There<br />
may be a leak in the system or process debris<br />
in the pump, or the pump motor may have<br />
failed. Inspect the pipelines and service the<br />
pump.<br />
Refer to the recommended actions for error<br />
number 411 in this table.<br />
Table 6-1 - Warning messages<br />