1.10 Electrical system<br />

Refer to Figures 1-3 and 1-4 which show schematic diagrams of the electrical and control<br />

systems of the iH80, iH600 and iH1000 systems.<br />

1.10.1 Introduction<br />

Refer to Figures 1-3 and 1-4. The iH system has an Electrics Box (1), Electronics Module (3), a<br />

Gas Module (2) and a Pump Display Terminal (4). The iH1000 system also has an inverter<br />

(50). Optionally, your iH system may have an iH Interface Module, an iH Communications<br />

Module, an HCMB pump and other accessories, such as an exhaust temperature sensor (30)<br />

and a Pump Display Module (10).<br />

All of these Modules communicate through a serial data network. Each Module on the<br />

network has a microprocessor, which controls the network communications and also contains<br />

the default setpoints for the sensors in or connected to the Modules (see Section 2).<br />

1.10.2Electrics box<br />

Refer to Figures 1-3 and 1-4. The external electrical supply (5) is connected to the electrical<br />

supply connector (6) on the Electrics Box (1). The Electrics Box switches the electrical supplies<br />

to the HCDP pump and to the HCMB pump (on an iH600 system). The Electrics Box also<br />

provides the electrical supplies to the other Modules in the iH system. The electrical supply<br />

to the HCMB1000 pump in an iH1000 system is provided by the inverter (50).<br />

The Electrics Box has contactors for the HCDP pump and the HCMB pump (on an<br />

iH600/iH1000 system), pump power and current consumption sensors and short-circuit<br />

protection. The Electrics Box also has a transformer and fuses for the 24 V control circuits<br />

and fuses for the following supplies: iH Interface Module (11) and d.c. electrical supply (47).<br />

Three thermal-type, self-resetting fuses in the Electrics Box protect the d.c. electrical supplies<br />

to the network.<br />

Refer to Figure 1-5 detail D. The rear panel of the Electrics Box has the following lamp and<br />

connections:<br />

Power on lamp (7)<br />

Tool Interface Module<br />

connector (8)<br />

LON Interface Connector (9)<br />

Electrical supply connector (10)<br />

This lamp is on when the electrical supply to the<br />

Electrics Box is on.<br />

Use this to connect an iH Tool Interface Module to the iH<br />

system (see Section 8.5.3); or to connect an external<br />

emergency stop circuit to the iH system (see Section<br />

3.15); or to connect a Disconnect Box to the iH system<br />

(see Section 3.21).<br />

Use this to connect the iH system to a network.<br />

Use this to connect your external electrical supply to the<br />

iH system: refer to Section 3.16.3.<br />