• Refer to Section 6 to determine the cause of the fault and to rectify the fault.<br />

6. Look at the pressure gauge in your inlet pipeline:<br />

• If the pressure is increasing, the HCDP pump-motor terminal-box (and/or if you have an<br />

iH1000 system, the HCMB pump-motor terminal-box) is incorrectly wired: immediately<br />

shut down the iH system, correct the electrical wiring(refer to Section 3.16.2), then repeat<br />

this procedure from Step 3 again.<br />

• If the pressure is decreasing, continue at Step 7.<br />

7. After you have commissioned the iH system:<br />

• If you want to continue to operate the iH system, refer to Section 5.<br />

• Otherwise, shut down the system,: refer to Section 5.10.<br />

3.20.2 Configure and switch on an iH600 system<br />

CAUTION<br />

Ensure that the booster start delay is correctly adjusted for your installation. If it is not, you<br />

may damage the iH when you start it up.<br />

CAUTION<br />

You can only switch on and configure an iH600 system if you have a Single Equipment<br />

Monitor (SEM) connected to the iH system. If you do not have SEM, do not use the<br />

following procedure; contact your supplier or BOC <strong>Edwards</strong> for advice.<br />

CAUTION<br />

Do not operate an iH600 system with the maximum inlet pressure for longer than the<br />

maximum time specified in Table 2-1. If you do, the HCMB pump may overheat<br />

Notes:<br />

You will need a stop-watch to do the following procedure, which assumes that you are familiar<br />

with the operation of the SEM; where necessary, refer to the manual supplied with the SEM.<br />

You can only adjust the iH booster start delay up to a maximum of 255 seconds. If your required<br />

delay is longer than 255 seconds, you must fit an iH600 Pressure Control Kit accessory: refer to<br />

Section 8.4.3.<br />

Operation of the iH system with high inlet pressure may result in shaft-seal pressure warnings<br />

(refer to Table 6-1): the operation of the iH system will not be affected, and you can ignore these<br />

warnings.<br />

3-30 iH Dry Pumping Systems