• Refer to Section 6 to determine the cause of the fault and to rectify the fault.<br />
6. Look at the pressure gauge in your inlet pipeline:<br />
• If the pressure is increasing, the HCDP pump-motor terminal-box (and/or if you have an<br />
iH1000 system, the HCMB pump-motor terminal-box) is incorrectly wired: immediately<br />
shut down the iH system, correct the electrical wiring(refer to Section 3.16.2), then repeat<br />
this procedure from Step 3 again.<br />
• If the pressure is decreasing, continue at Step 7.<br />
7. After you have commissioned the iH system:<br />
• If you want to continue to operate the iH system, refer to Section 5.<br />
• Otherwise, shut down the system,: refer to Section 5.10.<br />
3.20.2 Configure and switch on an iH600 system<br />
CAUTION<br />
Ensure that the booster start delay is correctly adjusted for your installation. If it is not, you<br />
may damage the iH when you start it up.<br />
CAUTION<br />
You can only switch on and configure an iH600 system if you have a Single Equipment<br />
Monitor (SEM) connected to the iH system. If you do not have SEM, do not use the<br />
following procedure; contact your supplier or BOC <strong>Edwards</strong> for advice.<br />
CAUTION<br />
Do not operate an iH600 system with the maximum inlet pressure for longer than the<br />
maximum time specified in Table 2-1. If you do, the HCMB pump may overheat<br />
Notes:<br />
You will need a stop-watch to do the following procedure, which assumes that you are familiar<br />
with the operation of the SEM; where necessary, refer to the manual supplied with the SEM.<br />
You can only adjust the iH booster start delay up to a maximum of 255 seconds. If your required<br />
delay is longer than 255 seconds, you must fit an iH600 Pressure Control Kit accessory: refer to<br />
Section 8.4.3.<br />
Operation of the iH system with high inlet pressure may result in shaft-seal pressure warnings<br />
(refer to Table 6-1): the operation of the iH system will not be affected, and you can ignore these<br />
warnings.<br />
3-30 iH Dry Pumping Systems