Warning message Advisory message Meaning/recommended action<br />

WARNING 1011 SEE MANUAL There is a phase imbalance in the electrical<br />

supply to the HCMB pump. Check the<br />

electrical supply and the connections to the iH<br />

system. If the electrical supply and<br />

connections are correct, the pump motor may<br />

be faulty.<br />

WARNING 1201 When started, no electrical current was drawn<br />

by the HCDP pump. Check whether the pump<br />

motor is open circuit.<br />

WARNING 1401<br />

SERVICE DUE (R)<br />

WARNING 1601<br />

SERVICE DUE (P)<br />

WARNING 1801<br />

SERVICE DUE (P)<br />

WARNING 2001<br />

SERVICE DUE (C)<br />

xxxxx HRS RUN<br />

SINCE SERVICE<br />

xxxxx HRS RUN<br />

SINCE SERVICE<br />

xxxxx CYCLES<br />

SINCE SERVICE<br />

xxxxx CYCLES<br />

SINCE SERVICE<br />

If not, the current sensor may have failed.<br />

The iH system has operated for xxxxx hours<br />

since it was last serviced. Service the iH<br />

system.<br />

The iH system has operated for xxxxx hours<br />

since it was last serviced. Service the iH<br />

system.<br />

The iH system has been operated for xxxxx<br />

process cycles since it was last serviced.<br />

Service the iH system.<br />

The iH system has been operated for xxxxx<br />

switch-on/off cycles since it was last serviced.<br />

Service the system.<br />

WARNING 3113 There is a fault in the ADC (analogue-to-digital<br />

convertor) in the Control SMA; the Control<br />

SMA must be replaced.<br />

WARNING 3501 (a) the gas module mass flow transducer has<br />

drifted outside the valid range allowed for<br />

zeroing, or (b) you cannot zero the transducer<br />

because the nitrogen purge is switched on or<br />

the HCDP pump is on.<br />

WARNING 3509<br />

TOTAL N2 FLOW LO<br />

WARNING 3511<br />

TOTAL N2 FLOW HI<br />

WARNING 3911<br />

EXHAUST PRESSURE<br />

FLOW LOW<br />

CHECK GAS MODULE<br />

FLOW HIGH<br />

CHECK GAS MODULE<br />

EXHAUST BLOCKED<br />

SERVICE PUMP<br />

The total nitrogen flow is too low. There may<br />

be a fault in the Gas Module.<br />

The total nitrogen flow is too high. There may<br />

be a fault in the Gas Module.<br />

The pressure in the exhaust pipeline is too<br />

high. There may be process debris or<br />

condensation in the exhaust pipeline, or a<br />

valve in the pipeline may be shut. You may<br />

have too many pumping systems connected to<br />

the exhaust pipeline.<br />

Table 6-1 - Warning messages (continued)<br />