User ManUal - FlowVision GmbH

User ManUal - FlowVision GmbH

User ManUal - FlowVision GmbH

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

FC100-CA | Flow MeterConfigurationGMBH5.10.1 Limit switches – switch-on/switch-off valueDepending on the configuration limit values 1 and 2 may be set either for flow velocity or mediumtemperature.The limit value may be set over the entire display range (-40 °C … 130 °C | 0 m/s … 99.99 m/s) andis always related to the display value.Limit switch up-date is by measuring rate, independent of the set measuring time.The hysteresis is determined by entering different switch-on and switch-off values. Its magnitudeshould be reasonably adjusted to current operating conditions.A specific definition of the operation (closed-current or open-circuit principle) may be dropped byseparately entering the switch-on and switch-off value of the limit switch, because the definition isdeducted from the switch-on and switch-off value.Example 1: Switch-on value lower than switch-off value0switch-on valueswitch-off valuemeasured value(flow/temperature)ONOFFhysteresisswitching condition ONswitching condition OFFfig. 41Example for ON:FC100-CA with relay outputs (option R2):• LIM1 - LIM1COM = closed/LIM1 - LIM1COM = openFC100-CA with transistor outputs (option T4):• LIM1E - LIM1C = switchedExample 2: Switch-on value higher than switch-off value0switch-off valueswitch-on valuemeasured value(flow/temperature)ONOFFhysteresisswitching condition ONswitching condition OFFfig. 4255