GMBHFlow Meter | FC100-CAConfiguration5.13 Quitting the configuration menuTo quit the configuration menu, the controller will check the data entered for plausibility.“CONFIG. OK!” is indicated when the data are correct. The menu may than be quitted by pressingM MODE.Errors found during the plausibility check are indicated in the following sequence of priority.Priority of entry errors in the CONFIGURATION menu:ERR. A-OUT FLOWOUT OF RANGE→analogue output – flow velocityflow analogue output outside measuring rangeERR. A-OUT FLOWZERO ≥ FSERR. A-OUT TEMP.OUT OF RANGEERR. A-OUT TEMP.ZERO ≥ FS→→→analogue output – flow velocityinitial value ≥ final valueanalogue output – medium temperaturetemperature analogue output outside measuringrangeanalogue output – medium temperatureinitial value ≥ final valueerror LS1ON = OFF→switch-on value for limit switch 1 equals switch-offvalue for limit switch 1error LS2ON = OFF→switch-on value for limit switch 2 equals switch-offvalue for limit switch 2The menu can only be quitted after correction of the error(s). To do this, return to the beginningof the configuration menu by pressing UP or DOWN and select the menu option with theincorrect entry for correction.58