.- The <strong>12</strong>61<br />
Calorimeter<br />
Operates in a compact, self-contained<br />
jacket with a built-in oxygen filling<br />
system.<br />
EFFECTIVE THERMAL JACKETING<br />
Outstanding thermal Jacketing IS provided In the<br />
<strong>12</strong>61 Calorimeter by a circulating water system which<br />
maintainS a forced flow around the sides and bottom<br />
of the bucket chamber and through the cover as well.<br />
The water In this system IS recirculated continuously<br />
by a bUilt-in, high capacity pump. Since all tests are<br />
run by Isoperlbol methods, the Jacket temperature<br />
remaIns constant dUring a run and no water<br />
additions are needed. Also, no water additions or<br />
waiting period are required at the end of a run since<br />
the jacket does not have to be recycled back to a<br />
fixed starting temperature. Jacket heating IS provided<br />
by a sealed ImmerSion heater operated by the<br />
calorimeter controller In conjunction with a bUilt-in<br />
thermistor probe.<br />
MOLDED FIBERGLASS AND<br />
BRASS CONSTRUCTION<br />
To ensure minimal heat leak and long service life,<br />
both the outer case and the water jacket are made of<br />
a high strength, compression molded, fiberglass<br />
reinforced polyester resin. These <strong>part</strong>s are<br />
assembled Into a sturdy, light weight structure which<br />
can be used wIth aCidic water supplies or In a<br />
corrosive laboratory atmosphere without damage to<br />
the Instrument. The Inner bucket chamber IS made of<br />
nickel brass for good corrosion resistance and<br />
proven heat transfer characteristics.<br />
A RELIABLE OXYGEN BOMB<br />
Each <strong>12</strong>61 Calorimeter IS equipped wIth a Parr<br />
1108 Oxygen Bomb In which samples can be burned<br />
completely and safely with full recovery of all<br />
combustion products. The bomb IS held In an oval<br />
bucket, covered with two liters of water. An Impeller<br />
driven by a motor mounted Within the cover produces<br />
excellent Circulation around the bomb.<br />
A BUILT-IN OXYGEN FILLING SYSTEM<br />
To Increase the speed, reliability and safety of the<br />
bomb filling operation, the <strong>12</strong>61 CalOrimeter has a<br />
bUIlt-in automatic system for filling the bomb WIth<br />
oxygen to the desired operating pressure ThiS IS<br />
Installed Within the calOrimeter, leaving only a short<br />
filling hose extending from the Jacket To fill the bomb<br />
the operator SImply slips a connectIOn fitting onto the<br />
bomb Inlet valve and pushes a button to start the<br />
filling sequence Filling then proceeds automatIcally<br />
at a controlled rate to a pre-set pressure. He then lifts<br />
the connector from the bomb and proceeds WIth the<br />
test. If. for any reason, the bomb pressure did not<br />
reach the desired level, an error message Will<br />
indicate a low charge The filling rate IS regulated so<br />
that oxygen enters the bomb at a Uniform, controlled<br />
rate, avOiding any sudden blast that might blow a<br />
powdered sample out of the combustion cup. Safety<br />
provisions are Included to prevent an aCCidental<br />
overcharge to the full pressure of the oxygen supply<br />
tank.<br />
PREaSE ELECTRONIC THERMONIETRY<br />
Temperatures are measured with a high preCISion<br />
electrOniC thermometer bUilt Into the calOrimeter,<br />
uSing a speCially deSigned thermistor network to<br />
obtain excellent linearity and high resolution In<br />
differential temperature measurements. ThiS custom<br />
deSigned system WIth ItS associated software<br />
provides a umform scale over a Wide range of<br />
operating temperatures so that the energy eqUIvalent<br />
(EE value) for the calOrimeter remains remarkably<br />
constant as compared With less stable systems uSing<br />
diodes or other temperature sensors ThiS stability IS<br />
confmned by many users who report that they find It<br />
necessary to recheck the energy eqUivalent only<br />
once or twice monthly unless a major component of<br />
the system has been changed.<br />
The temperature sensor IS sealed In a 118" dla<br />
stainless steel probe held In a fixed mounting on the<br />
calOrimeter cover, where It IS raised and lowered With<br />
the cover Measurements over the 20 0 to 40 0 C<br />
workmg range can be obtained With 0.0001 0 C<br />
resolution The absolute accuracy IS less precise<br />
(+1-a 1 0 C) but thiS does not affect differential<br />
measurements. Tailored software converts all<br />
readings to CelSIUS values Instead of ambIguous<br />
resistance umts No mercurial thermometers are<br />
required.