CO 2 flux measurements with EGM: A field guideCO 2 measurements from soil and tree collars once a monthEquipmentEGM (CO2 analyser)ChamberComputerWaterpro<strong>of</strong> bagUSB–RS port cableTemperature sensorSoil moisture sensorData sheets or notebookPencilsPlot mapThe EGM must not get wet (it will break)!Don’t measure in heavy rain.Transport the EGM and the computer in awaterpro<strong>of</strong> bag.StartingConnect the chamber to the EGM.Turn the Charge / Run switch <strong>of</strong> the EGM to Run andturn the On / Off switch to On.Warm-up, wait 5 to 10 mins (EGM temperature will go up to ~50°C).1REC 2SET 3CAL4DMP 5CLR 6CLKPress 1 (REC)DATA FITTING1LINEAR 2QUAD.?Press 1 (LINEAR)1DT:120 2DC: 503APPROX.TEMP. 25Press YSOIL RESP. DATARECORD 1ALL 2ENDPress 1 (ALL)V: 1171 A: 78V/A*100: 1491Press YPress Y79