3.2 SPECIFICATIONS--MSM400HousingExternal dimensionsCable GlandsWeightABS with polycarbonate lid, IP65256.5 wide, 236.7 high, 95.0 deep, including wall mounting brackets3x 16mm holes and plastic glands supplied3x 20mm holes and plastic glands supplied2 kgWall Mounting holes 6 off Diam 5.0mm (See Drawing Section 4.5)Power supply options 115 V a.c. (±15%) 50 / 60 Hz230 V a.c. (±15%) 50 / 60 Hzor 24 V d.c. (15 to 30 Volts)Power consumptionFuse (F1)Current OutputHARTRelay OutputsDC Power OutputSensor connectionsTrigger inputsa.c. 10VAd.c. 6W200 mA (T) 5 x 20mm0-20 or 4-20 mA selectable,maximum load 1KΩ (at 22mA)maximum applied voltage 48v d.cHART digital communications, Two HART internal test terminals provided.2x SPCO Relays, rated 5 Amps at 250 V a.c. Resistive24V DC for external sensors such as <strong>Mobrey</strong> ElectrosensorTerminals for <strong>Mobrey</strong> sensor Tx and Rx cables, each 2 cores and screenUnit accepts two 5V d.c.trigger input signals. 5V d.c. provided - compatiblewith <strong>Mobrey</strong> ElectrosensorTerminals Max. cable size 2.5mm 2Ambient temperature -30°C to 55°CMax AltitudeMax HumidityElectrical SafetyInstallation Category2000m95% RHConforms to EN61010-1Cat III 132V a.c. Max., Cat II 264V a.c. Max.Pollution Degree 2EMCComplies with EN61326 (Industrial level)10IP258