06.01.2013 Views

air data test set mps 30 digital portable user manual - CALIBRATION ...

air data test set mps 30 digital portable user manual - CALIBRATION ...

air data test set mps 30 digital portable user manual - CALIBRATION ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

5.0EQUIPMENT DESCRIPTION<br />

5.1Front Panel<br />

MPS <strong>30</strong> User’s Manual DMMC/0803/CM<br />

Two pressure ports (or more in case of Multiple Isolator) on the front panel as<br />

shown in Figure 5-1 provide connections to the <strong>air</strong>craft pressure ports. No. 9 is for<br />

the static and No. 10 is for the pitot port. The diameters of the two quick-disconnect<br />

couplers are different sizes to avoid wrong connections. As option the ports can be 4<br />

(or 2, or 3) static and 4 (or 2, or 3) pitot, connected to normally open valves that can<br />

be open/close through the HHRCU to verify leakage.<br />

Manual regulators (3 and 4) are used to vent the system in case of electrical<br />

power interruption. If the electrical power failure continues, <strong>manual</strong>ly open (slowly,<br />

counter-clockwise) regulators 3 and 4 together, be sure the “V” No. 3 does not cause<br />

an excessive <strong>air</strong>speed, and the “E” No. 4 does not cause an excessive altitude rate.<br />

If the ADTS has been <strong>manual</strong>ly vented before the startup, close the needle<br />

valves before switching-on.<br />

Connections to external pressure/vacuum generators must be tightly closed<br />

(fittings 5 and 6 of Figure 5-1).<br />

Fitting 5 is the positive pressure port, and may be used to apply a positive<br />

pressure, not exceeding 2,000 hPa (relative). This may be used if the internal<br />

pressure generator of the MPS <strong>30</strong> is not adequate for the pitot load.<br />

Fitting 6 is the vacuum port. It may also be used as a vacuum source to fasten<br />

special adapters with suckers to the <strong>air</strong>craft (sometimes used for the static port).<br />

This port may be used to connect an external vacuum pump to the MPS <strong>30</strong> to<br />

replace the internal one.<br />

Electrical connectors, switches, and warning lights:<br />

- Connection to the electrical power: No. 2.<br />

- Connection to the HHRCU: No. 7.<br />

- Connection for remote control RS232 (optional) No. 8. This connector may be<br />

used to control all the functions with an external PC with high level instructions.<br />

Powerful for use of MPS<strong>30</strong> in ATE systems.<br />

Issue Aug. 03 D. Marchiori PROPRIETARY INFORMATION - 12<br />

This material not to be disclosed or reproduced.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!