MPS <strong>30</strong> User’s Manual DMMC/0803/CM<br />

The ">" symbol will now flash in front of Asp. value, insert the Airspeed to be<br />

simulated at that point or confirm it by pressing " GO".<br />

Continue inserting the Altitude rate and Airspeed Rate values.<br />

Now the ">" symbol flashes in front of: "Wait".<br />

The value to be entered corresponds to the time (in seconds) that the MPS<strong>30</strong><br />

will wait before passing automatically to the next step. If 0 (zero) value is entered,<br />

the MPS<strong>30</strong> will not pass to the next step unless the ▲ (up arrow) is pressed. (Note<br />

that, also if a value is entered, it is always possible to pass to the next step by<br />

pressing ▲ or to the previous one by pressing the ▼ keys).<br />

After the wait time is <strong>set</strong> up, the ">" flashes in front of L0: this corresponds to the<br />

stabilization time for the leak check (i.e. the time – in seconds to stabilize the<br />

pressure before the leak is measured.). If zero value is inserted, the MPS<strong>30</strong> will not<br />

perform the stabilization time for the leak <strong>test</strong>.<br />

When the stabilization time is <strong>set</strong> up, the ">" flashes in front of L1: this is the<br />

measuring time for the leak <strong>test</strong> (i.e. the time in seconds to measure the leakage). If<br />

Zero value is entered, the MPS<strong>30</strong> will not perform the leakage measurement.<br />

When the MPS<strong>30</strong> is performing the leak <strong>test</strong>, the following menu appears:<br />

READ<br />

SET<br />

PRESS<br />

ALTITUDE AIRSPEED RATE<br />

172 100.0 0<br />

A. 500 100.0 0<br />

LEAK WAIT 72 sec<br />

Pr: 1 0.160 FAR43&LK<br />

Pt/Ps MACH UNITS<br />

READ<br />

Issue Aug. 03 D. Marchiori PROPRIETARY INFORMATION - 32<br />

This material not to be disclosed or reproduced.<br />

SET<br />

PRESS