4.2 Additional Features<br />

The following functions and features are provided:<br />

MPS <strong>30</strong> User’s Manual DMMC/0803/CM<br />

• Leakage <strong>test</strong>. Using the LEAK key of the HHRCU, leakage is<br />

automatically measured in the desired engineering units per minute. The<br />

condition before the leakage <strong>test</strong> is restored, by pushing the Resume key.<br />

This function, also running with the multiple Isolator part number MPS<br />

<strong>30</strong>EMI (Optional) can separately control up to four Static and four Pitot<br />

lines in various combinations.<br />

• Vent function. At the end of the instruments <strong>test</strong>, this function<br />

automatically brings the static and pitot lines back quickly but safely to the<br />

atmospheric pressure. When the safe condition is reached, the control<br />

unit LCD reads Test Set and can be disconnected.<br />

• Selectable units: with the SET key of the Control Unit<br />

- For altitude: feet and meters.<br />

- For pressure: in Hg, hPa, Pa, mmHg, and psi.<br />

- For speed: kt, m.p.h. and km/h.<br />

• Center line correction. To input the altitude difference between the <strong>test</strong> <strong>set</strong><br />

and unit under <strong>test</strong> (UUT).<br />

• Programmability of the <strong>test</strong>. The <strong>test</strong>s can be made <strong>manual</strong>ly by typing<br />

the desired altitude, <strong>air</strong>speed, and vertical speed values with the<br />

keyboard or in two other ways: with pre-defined steps, or with 25<br />

complete pre<strong>set</strong> <strong>test</strong> profiles.<br />

• Vacuum generation. May be used to fasten special adapters with suckers<br />

to the <strong>air</strong>craft.<br />

• Limit values at startup: The ADTS is easily programmable for limit (not to<br />

exceed) values of the altitude, speed, vertical speed, Mach number, and<br />

speed ratio. The MPS <strong>30</strong> is delivered with the following pre<strong>set</strong> values<br />

from the factory:<br />

- Minimum altitude: –1,500 ft.<br />

- Maximum altitude: 50,000 ft.<br />

- Maximum <strong>air</strong>speed: 450 kt.<br />

- Maximum vertical speed: 6,000 ft/min.<br />

- V/t (IAS ratio): <strong>30</strong>0 kt/min.<br />

- Max Mach no.: 1<br />

The <strong>user</strong> can change these values. The <strong>test</strong> <strong>set</strong> can be furnished, upon<br />

request, with different default values.<br />

- Units of measurement at startup: feet, knots, inHg.<br />

• Automatic Protection against excessive leakage in Pitot line, running also<br />

with large Pitot volumes.<br />