MPS <strong>30</strong> User’s Manual DMMC/0803/CM<br />
step Alt Asp Rate Rate Wait L0 L1<br />
A -1500 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
B 0 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
C 500 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
D 1000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
E 1500 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
F 2000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
G <strong>30</strong>00 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
H 4000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
I 6000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
J 8000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
K 10000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
L 12000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
M 14000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
N 16000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
O 18000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
P 20000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
Q 22000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
R 25000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
S <strong>30</strong>000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
T 35000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
U 40000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
V 45000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
W 50000 100 <strong>30</strong>00 <strong>30</strong>0 90 0 0<br />
TABLE 1<br />
To run a correct <strong>test</strong> profile, the Unit Under Test (UUT), must be <strong>test</strong>ed before<br />
for leakage, both for the static and for the Pitot lines.<br />
profile.<br />
Leakage <strong>test</strong> can be programmed also from inside the above programmed <strong>test</strong><br />
If the leak <strong>test</strong> must be performed within the same <strong>test</strong> profile, TABLE 1 changes<br />
in TABLE 2 – next page (to exclude the static pressure control it is necessary to<br />
enter the Altitude rate = zero).<br />
line.<br />
In this table, the two steps A & B are related to the leakage of the static and pitot<br />
As previously described, the measured leakage is shown on the HHRCU as RA<br />
(for the static) and RS (for the Pitot).<br />
Issue Aug. 03 D. Marchiori PROPRIETARY INFORMATION - 36<br />