8.5.2 GO/NO GO Leak Test Procedure<br />
MPS <strong>30</strong> User’s Manual DMMC/0803/CM<br />
An automatic and programmable (Go/No Go) Leak Test procedure is available,<br />
i.e. both the initial <strong>test</strong> values and the <strong>test</strong> features can be programmed, and the <strong>test</strong><br />
is automatic. When the <strong>test</strong> is over, it shows if leaks on the static and pitot circuits –<br />
compatible with the maximum acceptable values are found. This procedures starts<br />
with Fc key (RED SHIFT + 3). The menu shows the <strong>set</strong>ting requirements, default<br />
values are already shown, following the table here below:<br />
Parameter Description Measure Units Value<br />
Stab. Time Parameters Stabilization Time seconds 60<br />
Total Time Total time (reading + stabilization) seconds 120<br />
AllwdFt/m. Allowable Static Loss Ft/min. 50<br />
Allwdkt/m. Allowable Pitot Loss Kt/min. 1<br />
@ Feet Quote to reach to start <strong>test</strong> Ft 10000<br />
With Ft/m Choose the rate Ft/min. 4000<br />
@ Knots Speed to reach to start <strong>test</strong> Kt 150<br />
KnLKl Choose the rate Kt/min. 100<br />
The driven menu shows some default values, with such values the operator can<br />
reach pre-<strong>set</strong> speed and quote from which – after the <strong>set</strong> stabilization time – start the<br />
measurements of the average losses, in the remaining period (total time –<br />
stabilization time). Press RETURN/GO key if default values match the desired ones,<br />
or insert in each box the required numeric value, followed by the RETURN/GO key.<br />
The <strong>test</strong> will be carried out automatically, as it is programmed, at the end it will<br />
show two average leak values – stressing possible excesses vs the admitted leaks<br />
(GO/NO GO TEST). To exit from the <strong>test</strong> – at any moment – press RETURN/GO<br />
key.<br />
Issue Aug. 03 D. Marchiori PROPRIETARY INFORMATION - 26<br />