56 General Description<br />

English<br />

Behavior<br />

before use<br />

• Keep an external defibrillator, oxygen, intubation<br />

equipment and emergency drugs on hand.<br />

• Before use, <strong>Reocor</strong> S should be visually inspected<br />

for damages and dirt.<br />

• Never use a device that is damaged or shows<br />

abnormal behavior. Replace any cable that shows<br />

even slight damage.<br />

• Before using <strong>Reocor</strong> S, the patient cable or leads,<br />

the user should touch the patient to equalize electrical<br />

potentials.<br />

• It is strongly recommended to examine all set<br />

parameters before the leads are connected to<br />

<strong>Reocor</strong> S.<br />

• Even though <strong>Reocor</strong> S is protected from dripping<br />

water, the device and all plugs should be kept<br />

clean and dry.<br />

• <strong>Reocor</strong> S cannot be sterilized.<br />

Lead connection • The connections of <strong>Reocor</strong> S and the temporary<br />

pacing leads must be secured and checked regularly.<br />

• The patient cable must first be connected to<br />

<strong>Reocor</strong> S and then to the leads.<br />

• The temporary leads, to which the <strong>Reocor</strong> S is connected,<br />

represent a low-impedance conductor to<br />

the myocardium for electric current. Therefore<br />

line-powered devices that are operated in the<br />

patient's vicinity must be grounded in accordance<br />

with established guidelines.<br />

• When handling already implanted leads, their connector<br />

pins and metal contact surfaces must not<br />

touch or come into contact with electrically conductive<br />

or wet surfaces.<br />

• If the cable is disconnected from the <strong>Reocor</strong> S, it<br />

must be reconnected immediately and the security<br />

of the connection has to be examined.<br />

• When using unipolar leads, two unipolar leads<br />

must be used for effective pacing.