57 General Description<br />

Behavior<br />

during use<br />

Pacing with<br />

high rates<br />

• During use of <strong>Reocor</strong> S, the protective cover must<br />

be completely closed to prevent accidental resetting<br />

of the programmed parameters.<br />

• Secure <strong>Reocor</strong> S either horizontally on a non-slip<br />

surface or on the patient with an armband, or<br />

operate it from a hanging position on the infusion<br />

stand using the hanger on the back of the device.<br />

• <strong>Reocor</strong> S must not be worn directly on the skin.<br />

• During use of <strong>Reocor</strong> S, the heart rate of the<br />

patient must be monitored with an ECG monitor<br />

with alarm function.<br />

• In case of disturbances caused by electromagnetic<br />

interference (EMI), <strong>Reocor</strong> S will switch to operating<br />

mode S00 when certain limits are exceeded.<br />

• Pacing the heart with rates higher than 180 ppm<br />

over a long period of time can cause severe hemodynamic<br />

complications. Pacing with high rates<br />

should only be performed when continuous monitoring<br />

is ensured.<br />

Behavior after use • After a defibrillation or cauterization, the device<br />

should be subjected to a function test.<br />

• If the device will be stored for a long period of time,<br />

the battery should be removed to prevent damage<br />

due to leakage.<br />

• A damp cloth and mild soap can be used for cleaning.<br />

Strong cleaning agents or organic solvents<br />

should be avoided, as these can corrode the<br />

plastic housing.<br />

• Inspection and maintenance work should be<br />

performed according to page 79.<br />

Battery operation • Do not use rechargeable batteries. The service<br />

time of these batteries is difficult to estimate,<br />

making it possible to inadvertently exceed the<br />

ERI 1) time, resulting in sudden cessation of pacing.<br />

English<br />

1) <strong>Reocor</strong> S reminds you to replace the battery with the ERI signal (Low Battery LED flashes).