76 Pacing Modes and Parameters<br />

Pacing Modes and Parameters<br />

Pacing Modes<br />

English<br />

Mode S00<br />

Mode SSI<br />

Mode SST<br />

High-frequency<br />

pacing<br />

WARNING!<br />

There are three pacing modes available: S00, SSI, SST<br />

as well as high-frequency pacing (Burst).<br />

For disturbances caused by electromagnetic interference<br />

(EMI), after certain limits are exceeded<br />

<strong>Reocor</strong> S will switch to operating mode S00 for the<br />

duration of the disturbance.<br />

The pacemaker sends pulses with a constant rate. The<br />

pulses are asynchronous, i.e. not synchronized with<br />

the intrinsic heartbeats.<br />

This mode functions when connected to the ventricle<br />

as V00 and to the atrium as A00.<br />

The pacemaker inhibits pulses when intracardiac<br />

potentials are sensed. It sends pulses if no event is<br />

sensed within one of the intervals corresponding to<br />

the selected frequency.<br />

This mode functions when connected to the ventricle<br />

as VVI and when connected to the atrium as AAI.<br />

The triggered pacing mode SST corresponds to the<br />

pacing modes SSI with the following distinction: no<br />

pulse inhibition takes place upon sensing of an event<br />

outside of the refractory period, instead pulse delivery<br />

is carried out immediately in the respective chamber.<br />

The rate of the burst function can be selected with the<br />

control dial (10) from 60 ppm to 1000 ppm. <br />

This function is activated with two key buttons: First<br />

the Select burst key button (11) must be pressed and<br />

then, within two seconds, the Start burst key button<br />

(8) must be pressed. The pulse delivery then lasts as<br />

long as the Start burst key button is pressed.<br />

After a burst stimulation in the atrium, the ventricular<br />

blanking interval can prevent sensing of intrinsic signals<br />

and lead to asynchronous pacing in the ventricle.