83 Technical Safety<br />

Technical Safety<br />

WARNING!<br />

The external pacemaker <strong>Reocor</strong> S meets the international<br />

standards for the safety of electro-medical<br />

devices according to IEC 60601-1 and IEC 60601-1-2,<br />

as well as the international standard IEC 60601-2-31<br />

for temporary, external pacemakers.<br />

The following special features offer safety for the<br />

patient:<br />

• No metal parts that can be touched, according to<br />

the definition of the IEC.<br />

• The design meets the standard for the device class<br />

CF (cardiac floating) and is approved for direct<br />

treatment of the heart. The pacemaker complies<br />

with the requirements for defibrillation protection<br />

stipulated in the international standards.<br />

• The closed protective cover protects the pacemaker<br />

against dripping water.<br />

The temporary leads that are connected to <strong>Reocor</strong> S<br />

represent a low-impedance conductor to the myocardium<br />

for electric current. Therefore line-powered<br />

devices that are operated in the patient's vicinity must<br />

be grounded in accordance with established guidelines.<br />

The pacemaker must not be used in areas at risk for<br />

explosion.<br />

All additional maintenance work and repairs should<br />

only be performed by <strong>BIOTRONIK</strong>.<br />