60 General Description<br />

Visual and Acoustic Signals<br />

English<br />

• During the self-test after switching on <strong>Reocor</strong> S,<br />

all LEDs light up and brief acoustic signals can be<br />

heard. The self-test is finished after a few seconds.<br />

• If the self-test does not find any errors, the LEDs<br />

and warning signals turn off.<br />

• When the self-test finds a defect, all LEDs flash<br />

continuously and warning signals sound.<br />

• A required battery replacement is indicated by the<br />

flashing red Low battery LED.<br />

• The Sense (green) LED signals sensing of a P wave<br />

or R wave.<br />

• The Pace (yellow) LED signals pulse delivery.<br />

• The LEDs and acoustic signals also provide the<br />

following warnings during operation:<br />

Warning Meaning Error correction<br />

Acoustic signal for<br />

two seconds<br />

Fast sequence of<br />

sounds<br />

Acoustic signal and<br />

flashing of the Pace<br />

and Sense LEDs<br />

Low battery LED<br />

flashes.<br />

A pulse amplitude of 1 V or<br />

a rate of 180 ppm is programmed.<br />

Impedance outside of the<br />

permissible range<br />

High rate protection has<br />

been triggered;<br />

self-test failed.<br />

Check whether the set<br />

values are suitable for<br />

the patient.<br />

Check whether all connectors<br />

are securely<br />

plugged in.<br />

Check whether the<br />

leads have the desired<br />

position.<br />

Turn the device off and<br />

return it to <strong>BIOTRONIK</strong>.<br />

ERI has been reached. Replace the battery;<br />

and about 36 hours a) of<br />

a) When using the battery type MN 1604 Duracell ® , Procell ®<br />

service time remain.