20.02.2013 Views

Drug Targeting Organ-Specific Strategies

Drug Targeting Organ-Specific Strategies

Drug Targeting Organ-Specific Strategies

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

tion has been deteriorated. However, pulmonary obstructions usually restrict the expiratory<br />

performance rather than the inhalation manoeuvre (e.g. [102,103]). Therefore, maximal inspiratory<br />

pressure (MIP) values of asthmatic and COPD patients can be of the same order of<br />

magnitude as that of healthy subjects [104]. Only severe COPD patients may not be able to<br />

generate MIPs higher than 1.5–2.0 kPa.The effect of airflow resistance and clinical condition<br />

on the generated pressure drop across an external airflow resistance is summarized in Figure<br />

3.5 (data derived from reference [104]).The data show that in vitro testing of dry powder<br />

inhalers at only 4 kPa, as prescribed by various guidelines, is inadequate for the prediction of<br />

their performance in practice.<br />

Attained flow parameters during inhalation can either be calculated (Eq. 3.1) from recorded<br />

pressure drop curves or measured directly, using the equipment described in Section 3.9.1.<br />

Various studies report peak inspiratory flow rate (PIFR) values for healthy and diseased<br />

adult subjects calculated from data on inhalation without external resistance (so-called ‘control<br />

values’)(e.g. [22,100,105,106]). As for the attainable pressure drop, asthma, COPD and<br />

cystic fibrosis may decrease PIFR, depending on the severity of the disease. The addition of<br />

an external airflow resistance, such as a DPI strongly affects the PIFR.Within the range of resistances<br />

for marketed DPIs, the average PIFR at maximal inhalation by healthy adults may<br />

decrease from 159 l min –1 (for R = 0.015 kPa 0.5 min l –1 , equals the Rhone Poulenc Spinhaler)<br />

to only 62 l min –1 (for R = 0.040 kPa 0.5 min l –1 , equals the Astra Turbuhaler) [22].<br />

A maximal average flow rate of approximately 60 l min –1 through the Turbuhaler has also<br />

been reported for asthmatics (e.g. references [103,107,108]).<br />

Various flow parameters may be relevant to good DPI performance.The effect of PIFR on<br />

the in vitro dose of fine particles from DPIs has been the subject of many studies (e.g.<br />

[109–112]. The actual dose of the fine particle fraction produced varies with the ranges of<br />

flow rates applied. The flow rate being the measuring principle used for the study and definition<br />

of the fine fraction. Most studies show a strong increase in fine particle output with increasing<br />

PIFR for the ASTRA Turbuhaler [110–112]. For the Glaxo Diskus and<br />

Diskhaler, the effect of PIFR is less extreme: in some cases an more or less constant, but<br />

also a much lower output of approximately 15 to 25% of the label claim, has been obtained<br />

[109,110]. Some recent studies refer to the fact that the Pulmicort Turbuhaler also has a<br />

constant fine particle yield if the acceleration towards peak flow (flow increase rate: FIR) is<br />

high enough [113,114]. For a FIR > 8 l s –2 , the fine particle output is already maximal at flow<br />

rates above 40 l min –1 . For capsule inhalers, the inhalation time has to be long enough for all<br />

the particles to pass rapidly through the narrow holes pierced in the capsule ends before inhalation<br />

stops.<br />

3.9.1 Measurement of the Inspiratory Flow Curve<br />

3.9 Inspiratory Pressure and Relevant Flow Parameters 77<br />

Many different techniques are available for flow measurement and for recording of respiratory<br />

functions or flow parameters in particular (e.g. [115,116]). However, not all methods are<br />

appropriate for measurement of inhalation flows, either because they have low frequency responses<br />

or they influence the shape of the inspiratory flow curve by a large volume or by the<br />

inertia of the measuring instrument (e.g. rotameters). They may also interfere with the<br />

aerosol cloud from the inhalation device during drug deposition studies.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!