Draft <strong>SPIRE</strong> <strong>Design</strong> <strong>Description</strong> Document<br />

Figure 4-9 - Back plane connections to the photometer LIA cards. The spectrometer LIA boards are conceptually<br />

indentical.<br />

LIA Amplifiers: The Lock-in Amplifiers are used to read out the AC signals from the detectors and convert<br />

them into DC. Each bolometer has a dedicated amplifier. The amplifiers are grouped together in groups of<br />

32 on each LIA board. Figure 4-9 shows the interconnection between the LIA boards on the motherboard.<br />

The analogue output from each of the amplifiers is placed on a bus connected to the DAQ + I/F board.<br />

Each individual amplifier has;<br />

(i) a differential input pre-amplifer circuit that uses Analog Devices OP-400 operational amplifier;<br />

(ii) an Intersil, DG303A digital switch. This takes the pre-amplified AC signal from the bolometer and<br />

full wave rectifies it using the signal from the bias generators. The switching is signal is phase<br />

commandable to compensate for any signal phase change that occurs between the detector and the<br />

digital switch in order to maximize the output from the switch;<br />

(iii) an active post low pass filtering circuit that removes the bias frequency and bias harmonic<br />

components from the rectified signal. The low pass filter circuit uses two Analog Devices OP-400<br />

operational amplifiers.<br />

These three components are illustrated in Figure 3-26. There are a total of 12 LIA boards; with nine being<br />

used for the photometer and three used for the spectrometer.<br />

The DAQ + I/F boards: There are a total of<br />

In order to limit the mechanical properties of this unit it is now divided into two boxes which are<br />

respectively the DCU and the FCU including the MCU, the SCU and PSU.<br />

The SCU function is manifold: it is in charge of interfacing FPU sub-systems such as the<br />

cooler heaters, the calibrators and the thermometry sensors plus housekeeping parameters<br />

with the DPU through digital interfaces. This function covers the sensors biasing, the signal<br />

amplification and the digitisation of the Analog parameters. The SCU is also in charge of<br />

providing all the sub-systems with secondary power supply lines from the S/C 28 V interface<br />

by means of the PSU (a set of DC/DC converters). This PSU is associated to a supply current<br />