B.2.2.1 Auto Focus Measurement with Glass Plate Opened<br />

Perform the measurement as follows:<br />

3. Open the scanner’s book cradle completely, as shown on the screen.<br />

Note:<br />

The position of the book cradles is verified. If the book cradles are not in the<br />

lowest (home) position, a Retry button <strong>and</strong> a message are displayed on the<br />

screen instead of the Next Step button .<br />

4. Press the Next Step button. The measurement starts.<br />

Picture 56: Auto Focus measurement with open glass plate<br />

After the Autofocus function has completed, the results will be displayed. Values displayed<br />

in green indicate valid results. Any error will be shown in red text, followed by some<br />

explanatory remarks.<br />

Note: It is normal that the measurement will return different values each time the<br />

measurement is repeated. The lens motor has a very high resolution <strong>and</strong> the<br />

best focal point has to be found in the already large focal range, therefore a<br />

variation of 50 – 100 in values is normal.<br />

Page 48<br />

<strong>Setup</strong> <strong>and</strong> Assembly Manual