Device communication<br />
capable of automatic configuration with the use of additional measurement description<br />
registers that are supplied by the logger.<br />
Table 4 Modbus measurement registers<br />
Register Number Data Type Read/Write Name Data Map<br />
40013-40014 4 byte float Read Measurement Register 1 Data Log Channel 1<br />
40015-40016 4 byte float Read Measurement Register 2 Data Log Channel 2<br />
40017-40018 4 byte float Read Measurement Register 3 Data Log Channel 3<br />
40019-40020 4 byte float Read Measurement Register 4 Data Log Channel 4<br />
40021-40022 4 byte float Read Measurement Register 5 Data Log Channel 5<br />
40023-40024 4 byte float Read Measurement Register 6 Data Log Channel 6<br />
40025-40026 4 byte float Read Measurement Register 7 Data Log Channel 7<br />
40027-40028 4 byte float Read Measurement Register 8 Data Log Channel 8<br />
40029-40030 4 byte float Read Measurement Register 9 Data Log Channel 9<br />
40031-40032 4 byte float Read Measurement Register 10 Data Log Channel 10<br />
40033-40034 4 byte float Read Measurement Register 11 Data Log Channel 11<br />
40035-40036 4 byte float Read Measurement Register 12 Data Log Channel 12<br />
40037-40038 4 byte float Read Measurement Register 13 Data Log Channel 13<br />
40039-40040 4 byte float Read Measurement Register 14 Data Log Channel 14<br />
40041-40042 4 byte float Read Measurement Register 15 Data Log Channel 15<br />
40043-40044 4 byte float Read Measurement Register 16 Data Log Channel 16<br />
4.1.A.4.4 Description registers<br />
Two description registers, A and B, are associated with each individual Modbus<br />
measurement register (Table 5). The Modbus description registers allow third parties to<br />
develop external devices that have plug and play capabilities with the FL900 logger and<br />
minimize manual setup by the end user. Each description register can be queried by an<br />
external device to get more details about the measurement value.<br />
54<br />
Table 5 Modbus description registers<br />
Register number Data type Read/Write Name<br />
40213-40214 4-byte Unsigned Read Measurement 1 Description Register. 'A'<br />
40215-40216 4-byte Unsigned Read Measurement 1 Description Register. 'B'<br />
40217-40218 4-byte Unsigned Read Measurement 2 Description Register. 'A'<br />
40219-40220 4-byte Unsigned Read Measurement 2 Description Register. 'B'<br />
40221-40222 4-byte Unsigned Read Measurement 3 Description Register. 'A'<br />
40223-40224 4-byte Unsigned Read Measurement 3 Description Register. 'B'<br />
40225-40226 4-byte Unsigned Read Measurement 4 Description Register. 'A'<br />
40227-40228 4-byte Unsigned Read Measurement 4 Description Register. 'B'<br />
40229-40230 4-byte Unsigned Read Measurement 5 Description Register. 'A'<br />
40231-40232 4-byte Unsigned Read Measurement 5 Description Register. 'B'<br />
40233-40234 4-byte Unsigned Read Measurement 6 Description Register. 'A'<br />
40235-40236 4-byte Unsigned Read Measurement 6 Description Register. 'B'<br />
40237-40238 4-byte Unsigned Read Measurement 7 Description Register. 'A'