92 Part C______________________________________________________________________________________________________MP UNIT Input and Offset RangeGain Setting(X)Input Signal Range(+/- mV)Input Resolution(uV)Offset Range(+/- mV)Offset Resolution(uV)MP36/35 MP30 MP36/35 MP30 MP36/35 MP30 MP36 MP35 MP30and MP45and MP45 (10 bit mode) and MP45and MP45(MP36/45 only) 5 200 N/A 61.0 N/A 100 N/A 8 N/A N/A10 1000 N/A 30.5 N/A 100 N/A 8 150 N/A20 500 N/A 15.3 N/A 100 N/A 8 150 N/A50 200 N/A 6.1 N/A 100 N/A 8 150 N/A100 100 70* 3.0 195 10 10 0.8 15 78.1200 50 50 1.5 97.7 10 10 0.8 15 78.1500 20 20 .61 39.1 10 10 0.8 15 78.11000 10 10 .30 19.5 4 10 0.3 5.2 78.12000 5 5 .15 9.77 4 10 0.3 5.2 78.12500 N/A 4 N/A 7.81 N/A 10 N/A N/A 78.15000 2 2 .06 3.91 4 10 0.3 5.2 78.110,000 1 1 .03 1.95 4 10 0.3 5.2 78.120,000 0.5 N/A .015 N/A 4 10 0.3 5.2 78.125,000 N/A 0.4 N/A .781 N/A 10 N/A N/A 78.150,000 0.2 0.2 .006 .391 4 10 0.3 5.2 78.1Notes *For MP30 only: This is a special case limit of the voltage range only for a gain of 100.The actual range from unit to unit may vary by as much as 20%.To save as a Preset, change the Channel Preset name and then click the New Channel Preset button.Visit the online support center at www.biopac.com