464 Plug-In Assembly ConfigurationNOTESince mercury-wetted relays are used in the assembly, thearrows on the relay card must always point upward to the12 o'clock position. Do not operate the assembly with therelays more than 30 degrees from the vertical position. Afterinstalling the assembly in the 3497A or 3498A, gently tapthe assembly to settle the mercury to the bottom of therelays.Figure 128 shows how to install the assembly in the 3497A and 3498A. After takingthe steps shown in the WARNING, CAUTION and NOTE above, plug the assemblyinto an unused slot in the 3497A or 3498A and turn the locking key clockwise tolock the assembly in place.SLOTS ZO-29.EIC.DIGITALINPUT CARDTERMINAL CARDDIGITALINPUT CARDIERHINPLChRDFigure 128. Option 110 - Installation in 3497Al3498AArtisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com