7 Aug-In Assembly ConfigurationHow to Install the High Voltage Actuator AssemblyFigure 136 shows how to install the assembly in the 3497A and 3498A. After takingthe steps shown in the WARNING and CAUTION notes above, plug the assemblyinto an unused slot in the 3497A or 3498A and turn the locking key clockwise tolock the assembly in place.Before connecting loads to the actuator, you should consider three configuration requirementsfor the assembly: (1 ) Peak Voltage, Current and Volt-Amps; (2) SwitchingSpeed and (3) Contact Protection.Peak Voltage, Current and Volt-AmpsDo not connect loads which exceed the maximum ratings in the following table.I HIGH VOLTAGE ACTUATOR MAXIMUM RATINGS (I Voltage Current Volt-Amps IAC 250 VRMS 2A RMS 500 VAIChannelDC 48V 2A 60 VAIChannelArtisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com