Part C Special applications, functional checks, service, and order numbers Sect. 7.7ExampleFull-scale range Q 100% = 200 m³/h (Fct. 1.01)Meter size DN = 80 mm = 3 inch (Fct. 3.02)Current at Q 0% I 0% = 4 mAQ 100% I 100% = 20 mA } (Fct. 1.05)Pulses at Q 100% P 100% = 200 pulses/h (Fct. 1.06)Flow sensor constant GK = 3.571 (see instrument nameplate)Constant (V in m³)(t in h) K = 7074 (see table)(DN in mm)Calculation of ”X” and setting of ”Y”X =Q100% x K 200 x 7074GK x DN 2 = 3.571 x 80 x 80= 61.905Y = 80, setting of switch Y, see GS 8 A front panel(comes closest to the value of X and is smaller than X).Calculation of setpoint readings I and fI = I 0% +Y 40(IX 100% - I 0% ) = 4 mA +61.905(20mA - 4mA) = 14.3mADeviations are permissible between 14.1 and 14.6 mA (equivalent to ± 1.5 %).f =Y 40x PX 100% =61.905x pulses / h = 180.9 pulses/hDeviations are permissible between 178.2 and 183.6 pulses/h (equivalent to ± 1.5 %).If you need to return your flowmeter to KROHNE,please refer to the last-but-one page of these instructions.05/2003 <strong>IFC</strong> <strong>110</strong> F 79