Sect. 8.1 Part C Special applications, functional checks, service, and order numbers8 Service8.1 Replacing the power supply fuseSwitch off the power supply before opening the housing.1) Remove the cover from the terminal compartment (remove 2 screws).2) Unscrew the cap of the power supply fuse F.3) Replace fuse F5 or F6, type 5 x 20 G (Order No. see Sect. 9).F5: value for 100-230 V AC 0.8 A T, breaking capacity 1500 AF6: value for 24 V AC / DC 1.6 A T, breaking capacity 150 AF5/F6FPI/OPE/FEPSRSSSCS-<strong>EEx</strong>Wpower supply fuses, values are indicated aboveplug-in terminal for field current supply line, 4-pinplug-in terminal for outputs and inputs, 2x8-pinU-clamp terminal for connecting protective conductor PE or functional groundingconductor FEplug-in terminal for power supply line, 3-pinplug-in terminal for interface(s)U-clamp terminal for connecting the signal line shielding:signal line A: 2nd shield (7)signal line B: 3rd shield (11)plug-in terminal for electrode signal line, 5-pinSame as “S”, only available for hazardous duty versioninternal connection, may not be removed.I/ORS80 <strong>IFC</strong> <strong>110</strong> F 05/2003