Sect. 8.7 Part C Special applications, functional checks, service, and order numbersInputs/outputs PCB, I/OJumper X3 + X6X3 +active operationX6 ++ X3passive operation+ X6Jumper X4DC operation ≤ 0.2 AAC operation ≤ 0.1 A (factory setting)X1 plug-in terminals inside terminal compartment Small fuses TR 5, values and OrderX2 internal connection to motherboard No. see Section 9X4 jumper, change-over of AC/DC operation F1 Pof output A1, see Sect. 6.3 F2 A1X5 multipoint connector F3 A2IC 11 control program EPROM F4 D1IC 12 microprocessor F5 A2 terminalsX3/X6 jumper for active or passive operation F6 C1of current output F7 C2F8 E+F9 E–HART ® / RS 485 PCBF1X1IC3S3F2F3IC1IC18S4X1 internal connection to mother board F1, F2, F3 small fuses TR5X2 plug-in terminal inside terminal IC _ _ integrated circuitcompartment S3, S4 solder bridges86 <strong>IFC</strong> <strong>110</strong> F 05/2003