34 Calibration Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />
Connections<br />
1. Connect the blood pressure pneumatic tubing to the <strong>Spot</strong> <strong>LXi</strong> and to the test station.<br />
2. Connect the USB cable to the <strong>Spot</strong> <strong>LXi</strong> and the computer.<br />
3. Start the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> repair software on the computer.<br />
4. Connect the power supply to the <strong>Spot</strong> <strong>LXi</strong>.<br />
Figure 12. Connect Diagram<br />
Voltage Calibration<br />
Description Part Number Qty Source<br />
IBM compatible computer Windows 2000, XP,<br />
NT4<br />
* Only requires one of the pressure meters<br />
1. Follow the steps in “Connections”.<br />
2. Adjust the power supply to 5.5 Vdc ±0.1 Vdc and power on <strong>Spot</strong> <strong>LXi</strong>.<br />
3. Select Calibrate > Voltage in the repair software. The <strong>Spot</strong> <strong>LXi</strong> display window goes<br />
blank.<br />
4. Read the voltage on the digital multi-meter (DMM) connected to the power supply.<br />
5. Type the voltage reading in the Actual Voltage field and click the Calibration button.<br />
6. Click the Ok button to accept the inputs.<br />
7. Click the Ok button to answer the question in the text dialog box “Are you sure you<br />
want to make changes permanent?”<br />
8. Enter your initials in the Calibration Signature field to complete the voltage calibration.<br />