56 Disassembly and Repair Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />

3. Remove the four screws from the mounting plate and save the fish paper. This<br />

exposes the circuit board.<br />

4. Remove the three screws on the board (one on top and one on either side of the<br />

probe well cover) and lift the board out of the housing.<br />

5. Disconnect all connectors and replace with a new board.