7<br />
Technical Overview<br />
System Description<br />
Battery System<br />
Battery Charger<br />
The <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> automatically measures systolic and diastolic pressure (excluding<br />
neonates), pulse rate, temperature (oral, adult axillary, pediatric axillary, rectal, and ear),<br />
and pulse oximetry (SpO 2 ) as well as calculates Mean Arterial Pressure (MAP).<br />
Furthermore, <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> allows the entry of height, weight, respiration rate, and<br />
pain level. <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> also calculates Body Mass Index (BMI) following height and<br />
weight entry.<br />
The device is intended to be used by clinicians and medically qualified personnel. It is<br />
available for sale only upon the order of a physician or licensed health care provider.<br />
Refer to the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> Directions for Use manual for complete information.<br />
<strong>Spot</strong> <strong>LXi</strong> includes a 6 Volt, 6 amp-hour sealed lead-acid battery that supplies all power to<br />
the device. An external power supply charges the battery. The battery is always on-line<br />
(floated) rather than switching out when attached to the external power supply.<br />
<strong>Spot</strong> <strong>LXi</strong> uses a fuse and a shunt diode for protection against reverse battery connections.<br />
The fuse is located on the main board in the negative battery line and a reverse battery<br />
connection causes it to blow and require repair. The negative line location is more likely to<br />
short the battery (+) wire to ground than short the battery (-) wire to the VBUS node.<br />
Provides two levels of power to fast-charge or to float the battery at fixed voltages using<br />
software control. The fast-charge level rapidly brings the battery up to near-full capacity (7<br />
Volts) using software to not overcharge the battery. The float level, set per the<br />
manufacturer’s recommendation, slowly tops-off the battery and can stay on indefinitely.<br />
“CPU I/O” Power Supply<br />
Provides power to the I/O pins of the DragonBall MPU and the attached peripheral IC’s.<br />
The nominal voltage is set to a value of 3.15 Volts so that both the 3.0 Volt MPU and 3.3<br />
Volts peripherals can operate within specifications. <strong>Spot</strong> <strong>LXi</strong> uses buck switching topology<br />
for efficiency and an on-off power latch directly controls the power supply.<br />
CPU Core Power Supply<br />
Powers the MPU core only and operates at the recommended Dragonball voltage range<br />
of 1.8 +/- 0.1 Volts; 2.0 Volts absolute max.<br />
59