5<br />

Error Codes<br />

Troubleshooting<br />

This chapter provides helpful information in troubleshooting the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong>.<br />

The following tables of error codes provide a quick reference of the descriptions and<br />

probable causes of error codes.<br />

Table 10. General Error Codes<br />

Code Description Corrective Action<br />

C12 Device outside operating temperature<br />

range.<br />

Change ambient temperature.<br />

C13 Low battery level. Charge battery.<br />

E30 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E31 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E32 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E33 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E38 Date and time not set. Set date and time (see “Date/Time Menu Options” on<br />

page 25)<br />

E42 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E44 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

E45 Internal malfunction. Contact Technical <strong>Service</strong>.<br />

Table 11. Blood Pressure Error Codes<br />

Code Description Corrective Action<br />

C01 Blood pressure reading cancelled by user. Retake blood pressure reading.<br />

C02 Unable to release cuff pressure. Check tubing for kinks and connection integrity.<br />

C03 Inflation too quick. Check tubing and connections.<br />

C04 Air leak. Check blood pressure cuff and tubing connections.<br />

C05 Unable to determine blood pressure. Check connections; restrict patient movement.<br />

C06 Unable to determine blood pressure. Check connections; restrict patient movement.<br />

C07 Internal NIBP error. Device will power down.<br />

E10 Cuff pressure limits exceeded. Device will power down.<br />

E11 Cuff pressure duration exceeded. Device will power down.<br />

37