<strong>Service</strong> <strong>Manual</strong> Maintenance and <strong>Service</strong> 79<br />
Battery Replacement<br />
<strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />
Caution Only use the Welch Allyn 4500-84 lead acid battery. Using an incorrect<br />
battery will cause damage to the <strong>Spot</strong> <strong>LXi</strong> and void the warranty.<br />
Caution Do not break the shrinkwrap around the battery.<br />
If necessary, replace the internal battery after heavy use or the battery no longer charges.<br />
1. Power off the <strong>Spot</strong> <strong>LXi</strong> and disconnect the AC power transformer cord.<br />
2. Remove the four screws holding the battery door using a phillips-head screwdriver.<br />
Remove the battery door to expose the battery.<br />
3. Tip the <strong>Spot</strong> <strong>LXi</strong> backward and slide the battery out. Disconnect the one-way<br />
connector and then attach a new battery to the connector as shown. The one-way<br />
connector ends only connect one way. Do this as quickly as possibly to prevent loss<br />
of clock time.<br />
Figure 13. Battery Removal and One-way Connector<br />
One-way<br />
connector<br />
4. Slide the battery into the compartment as far as it will go. Push the connector down<br />
into the case next to the battery.<br />
5. Replace the battery door and tighten each of the four screws.<br />
6. Connect the AC power transformer to the <strong>Spot</strong> <strong>LXi</strong> and allow the new battery to<br />
charge for approximately 6 hours. The rechargeable batteries in the Braun<br />
ThermoScan PRO 4000 thermometer requires an additional 1 hour to charge and the<br />
rechargeable batteries in the external printer require an additional 4 hours to charge.<br />
You can use the <strong>Spot</strong> <strong>LXi</strong> during this charging period via the AC power cord.<br />
If <strong>Spot</strong> <strong>LXi</strong> displays the E38 error code after power up, set the date (see “Date/Time<br />
Menu Options” on page 25).<br />
The battery is a non-spillable lead-acid battery and must be recycled or<br />
disposed of properly according to national or local regulations.