05.08.2013 Views

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

A<br />

General Unit Test<br />

A-D Noise Test<br />

Leak Test<br />

Inflation Test<br />

Dump Test<br />

Repair Test Specifications<br />

This appendix refers to the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong> without pneumatics (tubing and cuff),<br />

temperature probe, SpO 2 probe, or main battery attached, unless otherwise noted.<br />

Use the Repair Software for the performing the tests on the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong>. The<br />

standard test voltage, unless otherwise stated is 6.5 Vdc (+0/-0.25 Vdc). Unless<br />

otherwise stated, all calibrated volumes are ±10cc of the stated volume.<br />

The A-D Noise Test is defined as the amount of noise on the <strong>Spot</strong> <strong>LXi</strong> A-D pressure<br />

channel; over a 1 second sample time while 0.0 mmHg is applied to the Device pressure<br />

port. The maximum limit is 0.050 mmHg.<br />

The Leak Test is defined as the amount of pressure drop that is recorded over a 15-second<br />

interval with a 100cc cylinder attached to the <strong>Spot</strong> <strong>LXi</strong> pressure port and that volume<br />

having a stabilized pressure of 250 mmHg. The limit is 5 mmHg maximum.<br />

The Inflation Test is defined as the amount of time the <strong>Spot</strong> <strong>LXi</strong> pump can inflate a 250 cc<br />

cylinder from 5 mmHg to 210 mHg. The limit is 7 seconds, maximum.<br />

The Dump Test is defined as the amount of time it takes <strong>Spot</strong> <strong>LXi</strong> to deflate a 500 cc<br />

cylinder from 260 mmHg to less than 15 mmHg. The limit is 10 seconds.<br />

Pneumatic Accuracy Test<br />

Enter the Internal Configuration Mode to perform this test (see “Internal Configuration<br />

Mode” on page 24).<br />

Perform an Auto-zero before starting this test (“Blood Pressure Functional Check”, Step 1<br />

on page 35 through Step 5 on page 35). The Pneumatic Accuracy test is defined as the<br />

comparison of <strong>Spot</strong> <strong>LXi</strong>’s pressure measurement and applied pressure at 0, 50, 150, and<br />

250 mmHg. The specification for each reading is within ±3 mmHg of the target pressure<br />

except for the 0 mmHg reading which is within ±1.0 mmHg.<br />

81

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!