38 Troubleshooting Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />
Table 11. Blood Pressure Error Codes<br />
Code Description Corrective Action<br />
E20 Internal NIBP error. Device will power down.<br />
Table 12. SureTemp Plus Temperature Error Codes<br />
Code Description Corrective Action<br />
C22 Temperature time limit exceeded. Remove probe from patient.<br />
E0.1 Probe heater error. Retake reading. If problem persists, replace probe.<br />
E0.2 Thermometer probe or device malfunction. Replace probe. If problem persists, contact Technical <strong>Service</strong>.<br />
E0.4 Probe is over temperature. If problem persists, contact Technical <strong>Service</strong>.<br />
E0.5 Unable to determine temperature. Retake reading. If problem persists, replace probe.<br />
E0.6 Probe data error. Retake reading. If problem persists, replace probe.<br />
E0.7 Broken thermometer probe. Replace probe.<br />
E0.8 Cannot read the probe’s configuration<br />
information.<br />
Contact Technical <strong>Service</strong>.<br />
E0.8 Temperature module data error. Contact Technical <strong>Service</strong> to return the device.<br />
E0.8 Cannot read the device’s Error Log. This problem will correct itself. If it persists, contact<br />
Technical <strong>Service</strong>.<br />
E0.9 Broken thermometer probe. Replace probe.<br />
E4.0 Internal temperature malfunction. Retake temperature. If problem persists, contact Technical<br />
<strong>Service</strong>.<br />
E4.1 Internal temperature malfunction. Retake temperature. If problem persists, contact Technical<br />
<strong>Service</strong>.<br />
E4.2 Internal temperature malfunction. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.3 Internal temperature malfunction. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.4 Temperature malfunction. Restart device. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.5 Temperature malfunction. Restart device. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.6 Temperature malfunction. Restart device. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.7 Cannot initialize thermometer. If problem persists, contact Technical <strong>Service</strong>.<br />
E4.8 Thermometer needs to be calibrated. Contact Technical <strong>Service</strong>.<br />
E4.9 Probe well missing or installed improperly. Reinstall probe well.<br />
E5.0 Temperature heater error. If problem persists, contact Technical <strong>Service</strong>.<br />
E5.2 Heater failsafe failure. If problem persists, contact Technical <strong>Service</strong>.<br />
A^! Device outside operating temperature<br />
range.<br />
Change ambient temperature.<br />
Av! Device outside operating temperature<br />
range.<br />
Change ambient temperature.<br />
b^ Internal temperature malfunction. Contact Technical <strong>Service</strong>.<br />
bv Internal temperature malfunction. Contact Technical <strong>Service</strong>.