05.08.2013 Views

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

Spot Vital Signs LXi Service Manual - Frank's Hospital Workshop

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

40 Troubleshooting Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />

6. Use the bulb to manually inflate the device to 250 mmHg and confirm <strong>Spot</strong> <strong>LXi</strong> meets<br />

the leak test specification (see “Leak Test” on page 81).<br />

Temperature Functional Check<br />

Use the 9600 Plus Caligration Tester to check the SureTemp Plus or Braun ThermoScan<br />

PRO 4000 thermometer accuracy (see “Temperature Functional Check” on page 29).<br />

Masimo SpO 2 Functional Check<br />

Use the Masimo Tester to check the functionality of the SpO 2 module of <strong>Spot</strong> <strong>LXi</strong> (see<br />

“Masimo SpO2 Functional Check” on page 32).<br />

Nellcor SpO 2 Functional Check<br />

Use a Nellcor approved simulator to check the functionality of the SpO 2 module of <strong>Spot</strong><br />

<strong>LXi</strong> (see “Nellcor SpO2 Functional Check” on page 32).<br />

Functional Testing Procedures<br />

Perform the functional testing procedures in this section before returning <strong>Spot</strong> <strong>LXi</strong> into<br />

service.<br />

Verify you are using a calibrated pressure meter with an accuracy of better than ±3 mmHg<br />

and that the calibration certificate is traceable to NIST.<br />

Subtract the rated accuracy of the pressure measurement standard from the ±3 mmHg<br />

rated accuracy of <strong>Spot</strong> <strong>LXi</strong>. This is the pass/fail criteria to determine if the device is within<br />

calibration. If the differences between <strong>Spot</strong> <strong>LXi</strong> and the pressure measurement standard<br />

are within the pass/fail criteria at all specified pressures then the device is within<br />

calibration.<br />

1. Follow the steps in “Connections” on page 34.<br />

2. Enter the “Internal Configuration Mode” on page 24. Use the Navigation and Select<br />

buttons to select Blood Pressure > BP Calibration Check.<br />

3. Connect to the 500 cc cylinder.<br />

4. Select Test > Calibration in the repair software. The dialog box displays the device<br />

manometer reading, battery reading, and valve and pump status.<br />

5. Use the Navigation and Select buttons to select Blood Pressure > BP Calibration<br />

Check.<br />

6. Verify the valve on <strong>Spot</strong> <strong>LXi</strong> is closed (the menu choice shows “Open Valve”).<br />

7. Use the hand bulb to verify <strong>Spot</strong> <strong>LXi</strong> is within calibration specification at the target<br />

pressures of 250±5 mmHg, 150 ±5 mmHg, 50 ±5 mmHg, and 0 mmHg with each<br />

reading within ±3 mmHg of the target pressure except for the 0 mmHg reading which<br />

is within ±1.0 mmHg.<br />

8. Set the power supply to 5.6 Vdc (+0.3/-0.0 Vdc). Verify the voltage reading meets the<br />

specification (see “Voltage Calibration” on page 82).<br />

After completing this test, return the power suppy to 6.5 Vdc (+0/-0.25 Vdc).

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!