<strong>Service</strong> <strong>Manual</strong> Repair Test Specifications 83<br />

Temperature Probe Test<br />

<strong>Spot</strong> <strong>LXi</strong> displays “ORL” after you remove the blue thermometer probe from the blue<br />

probe well.<br />

SpO 2 Option Requirements<br />

SpO 2 Functional Test<br />

These tests are only valid on <strong>Spot</strong> <strong>LXi</strong> models with the Masimo or Nellcor SpO 2 option.<br />

See “Masimo SpO2 Functional Check” or “Nellcor SpO2 Functional Check” on page 32<br />

for the functional test.<br />

SpO 2 Mode Current Test<br />

Fail Safe Test<br />

Over Pressure Test<br />

Over 15 mmHg<br />

The SpO 2 Mode Current Test is defined as the amount of current, less the Blank Mode<br />

Current, that is drawn through the battery terminals, placing <strong>Spot</strong> <strong>LXi</strong> in the Blank Mode,<br />

actuating the SpO 2 mode and applying any SpO 2 signal to the device. The maximum limit<br />

is 120 mA maximum.<br />

Verify that <strong>Spot</strong> <strong>LXi</strong> can detect over pressure on the pneumatic system between<br />

296.0 mmHg and 329 mmHg.<br />

Verify that <strong>Spot</strong> <strong>LXi</strong> can detect static pressure over 15 mmHg for 180 seconds.