Resetting the spectrometer<br />

at the start of data collection<br />

If you find that the interferogram peak in the live display tends to drift from its<br />

normal location, turn on Reset Bench At Start Of Collection. This repositions the<br />

peak before data collection starts.<br />

Starting data collection<br />

with an external trigger<br />

Diagnostic and<br />

alignment features<br />

Turn on Start Collection At External Trigger if you want data collection to begin<br />

automatically when a signal is received from another application. Use this option<br />

only in specialized experiments—such as gas chromatography FT-IR—in which<br />

data collection is synchronized with some external event. Special hardware is<br />

required.<br />

When you click the Diagnostic tab, special diagnostic and alignment features appear.<br />

At the left side of the dialog box is a live display of the detector signal. This<br />

display is similar to the live display provided on the Bench tab. See “Using the<br />

live display” for more information.<br />

Above the live display are several indicators that you can click to check the<br />

operation of spectrometer components. See “Checking spectrometer components”<br />

for more information.<br />

If you have a sealed and desiccated Nexus spectrometer, you can use the Check<br />

Desiccant button to check the performance of the desiccant. See “Checking the<br />

desiccant” for more information.<br />

If your spectrometer can be aligned automatically, you can use the Align button to<br />

perform an automatic alignment of the interferometer to maximize the detector<br />

signal. See “Aligning the spectrometer” for more information.<br />

If the interferogram peak has drifted from its normal location, you can click the<br />

Reset Bench button to reposition the peak.<br />

OMNIC User’s <strong>Guide</strong> 141