The units are described in the next sections.<br />

Kubelka-Munk<br />

Select Kubelka-Munk to convert % reflectance spectra collected using the diffuse<br />

reflection (DR) technique to Kubelka-Munk units (see the following example). Since<br />

these units are similar to absorbance units, the Kubelka-Munk format is useful for<br />

searching diffuse reflectance spectra against absorbance spectra in commercially<br />

available libraries. (Kubelka-Munk spectra are actually searched in absorbance.)<br />

Because of pathlength variation, in many cases the concentration of samples run<br />

using the diffuse reflection technique do not vary linearly with log (1/R) (which is<br />

the equivalent of absorbance for reflection measurements). The Kubelka-Munk scale<br />

was developed to provide a more linear relationship with respect to concentration.<br />

When you develop quant methods for determining component concentrations in<br />

sample mixtures, you can use spectra in log (1/R) or Kubelka-Munk units. Kubelka-<br />

Munk units are normally preferred because they eliminate any wavelengthdependent<br />

specular reflection effects, which reduce the accuracy of the analysis.<br />

However, the Kubelka-Munk model is accurate only if these criteria are met:<br />

• You are using high quality DR spectra.<br />

• The sample is diluted with a nonabsorbing matrix material such as KBr.<br />

• The sample concentration is low (about 1% sample to 99% matrix by weight).<br />

• The sample and matrix material are finely ground (a particle size of 2 to<br />

5 micrometers is recommended).<br />

• The mixture is uniform in particle size and composition.<br />

• The sample thickness must be at least 3 mm.<br />

188 Thermo Nicolet