When you click the button, the check begins and a gauge shows its progress:<br />

When the check is complete, the results appear:<br />

A green check mark indicates that the desiccant is still effective. Choose OK to<br />

close the dialog box.<br />

A yellow circle indicates that the desiccant is no longer effective and should be<br />

replaced or renewed. Click the Explain button for instructions.<br />

Note<br />

This feature measures humidity more accurately than the indicator under the<br />

desiccant compartment cover. For this reason you should rely on the Check<br />

Desiccant button to monitor the performance of the desiccant when the spectrometer<br />

is in use. Rely on the indicator only when the spectrometer is in storage. ▲<br />

Aligning the spectrometer<br />

Note<br />

Important<br />

If your spectrometer can be aligned automatically, you can click the Align button<br />

on the Diagnostic tab to perform an automatic alignment of the interferometer to<br />

maximize the detector signal.<br />

The spectrometer power should be on for at least 15 minutes (1 hour or longer for<br />

best results) before you perform an alignment. Aligning the system before it has<br />

warmed up and stabilized may give inconsistent results. ▲<br />

Set Gain to 1 before clicking the Align button. ▲<br />

The detector signal is shown in the live display during portions of the alignment.<br />

(The signal may take a few seconds to appear, depending on your spectrometer<br />

model.) Remove any accessories or samples from the sample compartment before<br />

aligning the spectrometer.<br />

Align the spectrometer’s interferometer after installing a new beamsplitter,<br />

replacing the laser or moving the spectrometer. It is also a good idea to align the<br />

interferometer if the signal intensity has dropped significantly from its usual level.<br />

Note<br />

Sampling accessories and some spectrometers must be aligned manually, not by<br />

using the Align button. If your accessory or spectrometer requires manual alignment,<br />

watch the live display on the Bench tab while making the appropriate adjustments on<br />

the accessory or spectrometer. The goal of manual alignment is to increase the<br />

maximum (Max) value above the live display to its highest possible level. See the<br />

manual that came with the accessory or spectrometer for instructions. See “Using the<br />

live display” for more information about the live display. ▲<br />

Configuring the system<br />

You can configure the system for the hardware you are using or perform other<br />

tasks described below by clicking the Configure Bench button. You must<br />

configure the system after you do any of the following:<br />

OMNIC User’s <strong>Guide</strong> 143