OPTIMIZED LASER PLATFORMS<br />

5.3.2. Ablation patterns on plastic<br />

Figure 5.3 shows pairs (top row for flat surfaces, <strong>and</strong> bottom row for spherical<br />

surfaces) of representative ablation patterns for all lasers of the study. In all cases, a<br />

correction of –9 D was programmed in the laser. The ablation patterns differed notably<br />

across lasers. While the Alcon <strong>and</strong> the Bausch & Lomb lasers provided highly<br />

symmetric patterns on both flat <strong>and</strong> spherical surfaces, the pattern delivered by the<br />

Wavelight laser showed asymmetries which were highly consistent across<br />

measurements.<br />

Fig. 5.3. a) Examples of ablation patterns obtained from ablated flat <strong>and</strong> spherical<br />

surfaces for the three different lasers, <strong>and</strong> a correction of -9 diopters. b) Horizontal<br />

sections of the ablation patterns on the flat surfaces (red lines), <strong>and</strong> other patterns<br />

obtained from different flat samples ablated on identical conditions (blue lines).<br />

Efficiency effects were responsible for a lower ablation depth in peripheral areas on<br />

spherical than on flat surfaces. This effect is notably larger in low fluence lasers (i.e.<br />

Bausch & Lomb laser) than higher fluence lasers (i.e. Alcon).<br />

Figure 5.3 (b) shows horizontal profiles of the ablation patterns on flat surfaces.<br />

The red lines correspond to the examples shown 2-dimensionally, while the blue lines<br />

correspond to different repetitions corresponding to different samples. The high<br />

similarity of the laser profiles estimated from different ablated samples (with the same<br />

laser <strong>and</strong> similar conditions) illustrates the high repeatability, particularly of the Alcon<br />

<strong>and</strong> Bausch & Lomb lasers. When asymmetries were present, as found with the<br />

Wavelight laser, the differences across repeated measurements also increased. The<br />

145